Defibtech, manufacture of the LifeLine and the ReviveR automated external defibrillators (AEDs), has issued a voluntary recall on 5,000+ DBP-2800 Battery Packs.
Defibtech has issued a voluntary recall on 5,000+ DBP-2800 Battery Packs that were manufactured prior to June 2007. The reason for the voluntary recall is due to the fact that the automated external defibrillator (AED) may falsely detect an error condition during the charging for a shock and it could cancel the shock therapy even though it may be needed.
Defibtech has secured the serial numbers for all the affected DBP-2800 Battery Packs manufactured prior to June 2007. To see if your DBP-2800 Battery Pack is affected, CLICK HERE.
If your DBP-2800 Battery Pack is one of the affected battery packs, please contact Defibtech for details about getting the battery pack issue corrected as soon as possible.
To view more details about the Defibtech Voluntary Recall, CLICK HERE.